G4 Multi Channel Force Instrument
9-5
Program Upgrade
Upgrading the program with a different version.
Upgrading is normally done from a USB memory. When upgrading the user is asked to
select the version to upgrade to by selecting an ‘Upgrade.txt’ file that is situated within
the folder containing all files and folders necessary for program upgrade. When
upgrading is done (a success message is displayed) the instrument will be restarted.
Note that it may take a few minutes for the instrument to restart after upgrading.
In case of failure when upgrading an error message will be displayed and the user will
be given a possibility to recover from the failed upgrade. Corrupt files or malfunctioning
USB memory are possible causes of upgrade failure.
Note that the folder containing the program upgrade must not be tampered with but
always handled as a whole. Any changes are likely to make the upgrade worthless and
making the instrument defective.
Note that several MB of free internal flash disk memory is required to be able to
perform a program upgrade. If the upgrade fails, due to file copy error, try to free some
space in the ‘user tree’ area by using the ‘File Handling’ maintenance function to
remove unnecessary files. It is also possible to use an ftp (File Transfer Protocol) client
to access the user area from an external computer over an Ethernet connection. See
chapter ‘Communication – Ftp Server’.
Instrument Restart
With the Instrument Restart function it is possible to force the instrument to make
complete restart (corresponding to a power down – power up cycle). This must be used
with great caution. The user must make sure that no hazardous situation will occur.
To use the Instrument Restart function requires either a Set-up Code or an Operator
Code if a lock is activated.
Note that the Instrument Restart function can be used even if a remote user has
exclusive access to parameter editing or maintenance functions. The purpose of this is
to be able to resolve a blocking between local and remote user.
When the user have selected the Instrument Restart function and entered a possible
Code he will be asked: “Restart the Instrument?” If pressing the Yes button a warning
message is displayed. Clicking Yes will make the instrument restart.
After some time, approximately one minute the instrument will be in operation again.
Содержание G4-DT
Страница 1: ...G4 Multi Channel Force Instrument Program version 101 1 0 0 Technical Manual PM DT HE types GB ...
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 36: ...Technical Manual 2 18 ...
Страница 45: ...G4 Multi Channel Force Instrument 4 5 ...
Страница 76: ...Technical Manual 4 36 ...
Страница 82: ...Technical Manual 5 6 ...
Страница 92: ...Technical Manual 6 10 ...
Страница 138: ...Technical Manual 7 46 ...
Страница 158: ...Technical Manual 8 20 ...
Страница 164: ...Technical Manual 9 6 ...
Страница 170: ......
Страница 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Страница 172: ...Technical Manual ...
Страница 173: ......