G4 Multi Channel Force Instrument
4-35
Range/Alternatives
Explanation and
<default value>
result of alternatives.
AOUT 1 Bandwidth (- AOUT 4 Bandwidth)
1.6 Hz
3 Hz
6 Hz
12 Hz
25 Hz
50 Hz
100 Hz
200 Hz
400 Hz
<100 Hz>
Sets the analog output bandwidth.
Note: This parameter is not shown if parameter ‘AOUT 1 Source’
is set to ‘Not in use’.
AOUT 1 Low Adjust (- AOUT 4 Low Adjust)
Range:
+/-2.000
Unit: %
<0.000>
Gives a possibility to adjust the offset of the analog output. This
allows for the reading of an external instrument connected to the
analog output to be fine-tuned. Full adjustment range
corresponds to approxi-2% of maximum analog output.
The parameter value will be set to zero each time
‘AOUT 1 Output Type’ is changed.
Note: This parameter is not shown if parameter ‘AOUT 1 Source’
is set to ‘Not in use’.
AOUT 1 High Adjust (- AOUT 4 High Adjust)
Range:
+/-2.000
Unit: %
<0.000>
Gives a possibility to adjust the gain of the analog output signal.
This allows for the reading of an external instrument connected
to the analog output to be fine-tuned. Full adjustment range
corresponds to approxi-2% of maximum analog output.
The parameter value will be set to zero each time
‘AOUT 1 Output Type’ is changed.
Note: This parameter is not shown if parameter ‘AOUT 1 Source’
is set to ‘Not in use’.
Содержание G4-DT
Страница 1: ...G4 Multi Channel Force Instrument Program version 101 1 0 0 Technical Manual PM DT HE types GB ...
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 36: ...Technical Manual 2 18 ...
Страница 45: ...G4 Multi Channel Force Instrument 4 5 ...
Страница 76: ...Technical Manual 4 36 ...
Страница 82: ...Technical Manual 5 6 ...
Страница 92: ...Technical Manual 6 10 ...
Страница 138: ...Technical Manual 7 46 ...
Страница 158: ...Technical Manual 8 20 ...
Страница 164: ...Technical Manual 9 6 ...
Страница 170: ......
Страница 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Страница 172: ...Technical Manual ...
Страница 173: ......