Technical Manual
2-10
AOUT1 and AOUT4
The voltage levels on connectors of I/O modules shall not exceed
hazardous voltage levels of 30 Vrms, 42.4 Vpeak or 60 Vdc under normal
conditions. In wet locations these voltage levels shall not exceed 16 Vrms,
22.6 Vpeak or 35 Vdc.
Analog output unit
The AOUT4 unit has 4 analog output channels, independently
isolated by operational insulation.
The AOUT1 has 1 output channel, isolated by operational
insulation.
The analog output signal will be connected to terminals
10, 11 (channel 1)
7, 8 (channel 2)
4, 5 (channel 3)
1, 2 (channel 4).
Shielded cable/cables must be used and the shield(s) must be
connected to the shield terminal(s) 12, 9, 6 and 3.
The shields can be connected to the most convenient ground/earth
point.
OUT
0V
Shield
AOUT1 Channel 1
10
11
12
+
OUT
0V
Shield
Channel 1
10
11
12
OUT
0V
Shield
Channel 2
7
8
9
OUT
0V
Shield
Channel 3
4
5
6
OUT
0V
Shield
AOUT4 Channel 4
1
2
3
+
+
+
+
Содержание G4-DT
Страница 1: ...G4 Multi Channel Force Instrument Program version 101 1 0 0 Technical Manual PM DT HE types GB ...
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 36: ...Technical Manual 2 18 ...
Страница 45: ...G4 Multi Channel Force Instrument 4 5 ...
Страница 76: ...Technical Manual 4 36 ...
Страница 82: ...Technical Manual 5 6 ...
Страница 92: ...Technical Manual 6 10 ...
Страница 138: ...Technical Manual 7 46 ...
Страница 158: ...Technical Manual 8 20 ...
Страница 164: ...Technical Manual 9 6 ...
Страница 170: ......
Страница 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Страница 172: ...Technical Manual ...
Страница 173: ......