Technical Manual
4-26
Range/Alternatives
Explanation and
<default value>
result of alternatives.
Menu ‘Fieldbus’
Note: The Fieldbus menu is not shown if parameter ‘Fieldbus‘ (in Hardware Config
menu) is set to ‘Not In Use’. See chapter 7 ‘Communication
−
Fieldbus interface’ for
more details on fieldbus configuration and usage.
Address
Range 1 - 125
< 126 >
Profibus address setting.
Note: This parameter is only shown if parameter ‘Fieldbus‘ (in
Hardware Config menu) is set to ‘Profibus’.
Address
Range 0 - 63
< 63 >
DeviceNet address setting
Note: This parameter is only shown if parameter ‘Fieldbus‘ (in
Hardware Config menu) is set to ‘DeviceNet’.
Baudrate
125 kbps
250 kbps
500 kbps
Auto
< Auto >
Sets the baudrate for a DeviceNet type of module. Must be
selected to suite the master.
125 kbps: Fixed baudrate 125 kbits/s.
250 kbps: Fixed baudrate 250 kbits/s.
500 kbps: Fixed baudrate 500 kbits/s.
Auto: Auto setting 125 – 500 kbits/s.
Note: This parameter is only shown if parameter ‘Fieldbus‘ (in
Hardware Config menu) is set to ‘DeviceNet’.
No of Data Blocks
Range 0 - 12
< 0 >
Sets the no of Data Blocks that should be mapped to the
memory available from the fieldbus. Each Data Block is
configurable.
Содержание G4-DT
Страница 1: ...G4 Multi Channel Force Instrument Program version 101 1 0 0 Technical Manual PM DT HE types GB ...
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 36: ...Technical Manual 2 18 ...
Страница 45: ...G4 Multi Channel Force Instrument 4 5 ...
Страница 76: ...Technical Manual 4 36 ...
Страница 82: ...Technical Manual 5 6 ...
Страница 92: ...Technical Manual 6 10 ...
Страница 138: ...Technical Manual 7 46 ...
Страница 158: ...Technical Manual 8 20 ...
Страница 164: ...Technical Manual 9 6 ...
Страница 170: ......
Страница 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Страница 172: ...Technical Manual ...
Страница 173: ......