Technical Manual
4-18
Range/Alternatives
Explanation and
<default value>
result of alternatives.
1:HSWF Update Rate
25 Hz
50 Hz
100 Hz
200 Hz
400 Hz
800 Hz
800 Hz ++
<800 Hz>
Defines the update rate for the measuring channel belonging to
Function Block 1 if the force conversion module is of HS WF 2
type.
Note: this parameter is only shown if the force conversion
module belonging to this Function Block 1 is of HS WF2 type.
1:WFIN Update Rate
9.3 Hz
19 Hz
37 Hz
75 Hz
150 Hz
300 Hz
300 Hz ++
<300 Hz>
Defines the update rate for the measuring channel belonging to
Function Block 1 if the force conversion module is of WF IN or
WF IN2 type.
Note: this parameter is only shown if the force conversion
module belonging to Function Block 1 is of WF IN or WF IN2
type.
1:Calibration Type
Known Force
Table
< Table >
Defines the type of calibration to be performed. A new
calibration is initiated as a ‘Calibration type’ is selected and
Enter key pressed.
Known Force: Known Force calibration is normally
the most accurate calibration type. It is recommended to load
with at least 2/3 of the used range.
Table: Table calibration is used when calibrating with data from
transducer calibration data sheet or when entering recorded
values from a previous calibration into a replacement instrument.
Содержание G4-DT
Страница 1: ...G4 Multi Channel Force Instrument Program version 101 1 0 0 Technical Manual PM DT HE types GB ...
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 36: ...Technical Manual 2 18 ...
Страница 45: ...G4 Multi Channel Force Instrument 4 5 ...
Страница 76: ...Technical Manual 4 36 ...
Страница 82: ...Technical Manual 5 6 ...
Страница 92: ...Technical Manual 6 10 ...
Страница 138: ...Technical Manual 7 46 ...
Страница 158: ...Technical Manual 8 20 ...
Страница 164: ...Technical Manual 9 6 ...
Страница 170: ......
Страница 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Страница 172: ...Technical Manual ...
Страница 173: ......