
Technical Manual
2-6
AC SUPPLY – 110-240 V
~
Power supply unit
Mains supply cable shall be separated and routed away from SELV
or SELV-E field wiring.
Remove power before removing the module from the instrument.
Make sure the fixation screws are well tightened during operation.
110-240 V~ In
Terminals 1, 2 and 3.
The AC SUPPLY gives the possibility to power the instrument from
110-240 VAC, connected according to the diagram below.
To achieve protective grounding, terminal 3 should be connected
to protective earth.
See Technical data for input voltage ratings.
24 V
Out
Terminals 4 and 5 can be used to supply max. 100 mA to digital
inputs and outputs.
Connection should be according to the diagram below.
See technical data for ratings.
1
AC SUPPLY
2
3
Shield
4
5
0 V
6
Shield
Protective earth
0 V
110-240 VAC
+24 VDC Out
+24 V Out
110-240 V~
Содержание G4-DT
Страница 1: ...G4 Multi Channel Force Instrument Program version 101 1 0 0 Technical Manual PM DT HE types GB ...
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 36: ...Technical Manual 2 18 ...
Страница 45: ...G4 Multi Channel Force Instrument 4 5 ...
Страница 76: ...Technical Manual 4 36 ...
Страница 82: ...Technical Manual 5 6 ...
Страница 92: ...Technical Manual 6 10 ...
Страница 138: ...Technical Manual 7 46 ...
Страница 158: ...Technical Manual 8 20 ...
Страница 164: ...Technical Manual 9 6 ...
Страница 170: ......
Страница 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Страница 172: ...Technical Manual ...
Страница 173: ......