Technical Manual
2-8
10
15
13
12
11
14
16
Exc.-
Exc.+
Sign.+
Sign.-
4-wire connection
Transducer
Transducer input, ch. 2
Exc+
Sense+
Exc-
Sense-
Sign+
Sign-
Shield
Junction box
17
22
20
19
18
21
23
Transducer input, ch. 1
Exc.-
Exc.+
Sign.+
Sign.-
Transducer
Transducer
Exc.-
Exc.+
Sign.+
Sign.-
6-wire connection
Exc+
Sense+
Exc-
Sense-
Sign+
Sign-
Shield
A transducer may be connected directly to terminals at the transducer input.
For several transducers or long distance, a junction box
and lengthening cable is needed.
For a HS WF2 channel, the shield can be connected to
ground/earth at any point