G4 Multi Channel Force Instrument
10-5
Set-up errors
These errors occur only during instrument set-up from the front panel.
Certain errors depend on more than one set-up parameter and it is the operator’s
responsibility to locate and correct all faulty set-up parameters.
Error
code
Explanation
151
Input signal error
The input signal is not valid during calibration. Check the live force and input
signal at the bottom of the display in calibration menus.
152
Unstable transducer signal used
This is not an error but only a warning. The force was not stable during
calibration which might reduce the accuracy of the calibration. Wait a little
longer and try again. If the problem persist, check the mechanical and
electrical stability of the installation, check the live force and input signal at
the bottom of the display in calibration menus.
155
Illegal calibration direction
All forces with corresponding mV/V values must be increasing for increasing
calibration point number.
156
Calibration not finished (all calibration points not stored)
This warning is given if the user is leaving the calibration menu when a
known force calibration is started but not all calibration points are stored.
The user will have the possibility to continue calibration, leave the menu and
discarding all calibration changes or leave the calibration menu and keeping
the calibration changes done so far. Note that leaving an unfinished
calibration might give a faulty calibration of the Function Block.
192
Too high transducer signal in calibration point 2
The mV/V signal in calibration point two is too high (often due to a previous,
strange calibration).
194
Illegal calibration direction.
All forces with corresponding mV/V values must be increasing for increasing
calibration point number.
195
Illegal I/O module configuration
The I/O-module set-up (Hardware Configuration menu) is not consistent with
the allowed number of modules of the different types or the allowed
combinations of modules.
197
Illegal Function Block input configuration
Each transducer input must only be used once.
See menu ‘Hardware Configuration’.
198
Illegal Function Block input configuration
None existing transducer input used. See menu ‘Hardware Configuration’.
199
Illegal Function Block input configuration
No transducer input selected for a Function Block (each Function Block
must have at least one transducer input assigned).
See menu ‘Hardware Configuration’.
Содержание G4-DT
Страница 1: ...G4 Multi Channel Force Instrument Program version 101 1 0 0 Technical Manual PM DT HE types GB ...
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 36: ...Technical Manual 2 18 ...
Страница 45: ...G4 Multi Channel Force Instrument 4 5 ...
Страница 76: ...Technical Manual 4 36 ...
Страница 82: ...Technical Manual 5 6 ...
Страница 92: ...Technical Manual 6 10 ...
Страница 138: ...Technical Manual 7 46 ...
Страница 158: ...Technical Manual 8 20 ...
Страница 164: ...Technical Manual 9 6 ...
Страница 170: ......
Страница 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Страница 172: ...Technical Manual ...
Страница 173: ......