G4 Multi Channel Force Instrument
2-5
DC SUPPLY 24 V
Power supply unit
The output of the external DC source must be rated 24 V
, ±15%
including fluctuations, min. 40 W. The DC source must provide Double
Insulation between Mains parts and 24 V SELV or SELV-E Circuit,
and a limited-energy circuit (maximum available current of 8 A).
For the US market this energy limit can be achieved with an
ANSI/UL248-14 fuse rated 5A.
For other markets an IEC 60127 T type fuse rated 4A may also be used.
24 V
In
Terminals 1, 2 and 3.
The G4 instrument should be powered by 24 V
,
connected according to the diagram below.
To achieve functional grounding, terminal 3
should be connected to ground.
See Technical data for input voltage ratings.
24 V
Out
Terminals 4 and 5 can be used to supply max. 100 mA
to digital inputs and outputs.
Connection should be according to the diagram below.
See technical data for ratings.
1
DC SUPPLY
2
0 V
3
Shield
4
5
0 V
6
Shield
0 V
Ground
0 V
+24 V In
+24 VDC In
+24 VDC Out
+24 V Out
Содержание G4-DT
Страница 1: ...G4 Multi Channel Force Instrument Program version 101 1 0 0 Technical Manual PM DT HE types GB ...
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 36: ...Technical Manual 2 18 ...
Страница 45: ...G4 Multi Channel Force Instrument 4 5 ...
Страница 76: ...Technical Manual 4 36 ...
Страница 82: ...Technical Manual 5 6 ...
Страница 92: ...Technical Manual 6 10 ...
Страница 138: ...Technical Manual 7 46 ...
Страница 158: ...Technical Manual 8 20 ...
Страница 164: ...Technical Manual 9 6 ...
Страница 170: ......
Страница 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Страница 172: ...Technical Manual ...
Страница 173: ......