Technical Manual
4-22
Range/Alternatives
Explanation and
<default value>
result of alternatives.
1:D Value Cal. P1
Range: +/–999999
Unit:
Measurement Unit
<0>
This parameter defines, for Input D, the load in the lowest
calibration point, normally 0.
Note: This parameter is only shown if parameter
‘Func. Block 1 Type’ is set to ‘4 Channel’ and ‘F.Block 1 Inp.D’ is
not set to ‘Not In Use’.
1:D Transd. Signal P1
Range: +/–9.99999
Unit: mV/V
<0.00000>
This parameter defines the transducer signal, for Input D, in the
lowest calibration point.
Note: This parameter value is automatically acquired, and
cannot be edited, if ‘1:Calibration Type’ is set to ‘Known Force’.
It is only shown if parameter ‘Func. Block 1 Type’ is set to
‘4 Channel’ and ‘F.Block 1 Inp.D’ is not set to ‘Not In Use’.
1:D Value Cal. P2
Range: +/–999999
Unit:
Measurement Unit
<1000>
This parameter defines, for Input D, the load in the second
calibration
point.
Note: This parameter is only shown if parameter
‘Func. Block 1 Type’ is set to ‘4 Channel’ and ‘F.Block 1 Inp.D’ is
not set to ‘Not In Use’.
1:D Transd. Signal P2
Range: +/–9.99999
Unit: mV/V
<2.00000>
This parameter defines the transducer signal, for Input D, in the
second calibration point.
Note: This parameter value is automatically acquired, and
cannot be edited, if ‘1:Calibration Type’ is set to ‘Known Force’.
It is only shown if parameter ‘Func. Block 1 Type’ is set to
‘4 Channel’ and ‘F.Block 1 Inp.D’ is not set to ‘Not In Use’.
1:D Set Zero
Range +/-999999
Unit:
Measurement Unit
< 0 >
This parameter is used to zero input D for Function Block 1.
Enter wanted value for the actual force on Input D, usually ’0’,
i.e. unloaded transducer.
Note: This parameter is only shown if parameter
‘Func. Block 1 Type’ is set to ‘4 Channel’ and ‘F.Block 1 Inp.D’ is
not set to ‘Not In Use’.
1:D Zero Offset
Range +/-999999
Unit:
Measurement Unit
< 0 >
This parameter shows the offset value acquired, for Input D, by
zeroing with ‘1:D Set Zero’. If this parameter is edited, the
zeroing of Input D will be influenced.
Note: This parameter is only shown if parameter
‘Func. Block 1 Type’ is set to ‘4 Channel’ and ‘F.Block 1 Inp.D’ is
not set to ‘Not In Use’.
Содержание G4-DT
Страница 1: ...G4 Multi Channel Force Instrument Program version 101 1 0 0 Technical Manual PM DT HE types GB ...
Страница 2: ......
Страница 18: ...Technical Manual 1 14 ...
Страница 36: ...Technical Manual 2 18 ...
Страница 45: ...G4 Multi Channel Force Instrument 4 5 ...
Страница 76: ...Technical Manual 4 36 ...
Страница 82: ...Technical Manual 5 6 ...
Страница 92: ...Technical Manual 6 10 ...
Страница 138: ...Technical Manual 7 46 ...
Страница 158: ...Technical Manual 8 20 ...
Страница 164: ...Technical Manual 9 6 ...
Страница 170: ......
Страница 171: ...G4 Multi Channel Force Instrument Appendix 1 ...
Страница 172: ...Technical Manual ...
Страница 173: ......