Parameters
130
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
Factor 2
Enters a factor to convert a datum to a result
Abs(WL2) x Factor = Result
(Abs Ratio, Abs Diff, Multiwavelength tests)
Factor 3-31
Enters a factor to convert a datum to a result
Abs (WL3-31) x Factor = Result
(Multiwavelength tests)
Graph
Displays the graph of collected data
(Kinetics, Scanning)
ID #
Enters the numeric identifier for measurement;
autoincrements during the test until turned off (set to 0)
Instrument Serial Number
Displays the serial number of the instrument
(Utility)
Intercept
Enters where the line crosses the Y-axis (Abs where conc=0)
Interval
Enters the wavelength range between data points
(Scanning tests only)
Interval Time
Enters the time between repeated readings
(Kinetics only)
Linearity Value
Enters a linearity value
(Kinetics only)
To help determine the linearity of the reaction during the
measurement, the instrument offers a linearity parameter. This
is the difference between the changes in the absorbance of two
measurements as shown in the following example:
Time
Abs
?A
Linearity
1
.1
--
---
2
.2
.1
---
3
.29
.09
P
4
.38
.09
P
5
.46
.08
P
6
.52
.06
F
Linearity is the ?A between ?A calculations.
P=Pass and F=Fail
Parameter
Description