Water Analysis Application
Using the One Point Adjustment Calibration Feature
Thermo Scientific
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
43
3. Prepare a blank, reagent blank (reverse color methods only) and calibration standard with
a known concentration within the method range and near the expected sample
concentration.
Refer to the specific reagent chemistry section for detailed instructions.
4. Press the
Calculate 1pt Adjust
function key.
The screen below will appear with the Standard Concentration: value highlighted.
5. Press
enter
to edit this field, then use the numeric keypad to enter the concentration of
the standard and press
enter
.
6. Wipe the exterior of the blank vial and place the blank vial into the holder in the sample
chamber.
7. Close the sample chamber door.
8. Press the
Measure Blank
function key.