5
Cell Correction
Cell Correction
Thermo Scientific
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
23
7. Press
Run Corr.
to start Cell Correction.
If you selected Discrete nms mode, first specify the wavelengths using the following
procedures, and then Cell Correction.
Cell Correction measures the other cells against the blank and records, stores and dates
the measurements. From these measurements Cell Correction establishes the required
correction factors, which then are automatically applied during all subsequent tests (if
Cell Correction is activated).
Specifying Wavelengths for Discrete nms Mode
To specify wavelengths for Discrete nms mode
1. Highlight
Sample Positioner
and press
Enter
to set this parameter to either Auto 3
(when using three large cell holders) or Auto 6 (when using six small cell holders).
2. Highlight
Number of Matched Cuvettes
and press
Enter
.
3. Specify the number of cells you are matching and press
Enter
.
4. Press
Set nms
to select the wavelengths for which Cell Correction will be run.
A list of wavelengths appears.