Water Analysis Application
Running a Water Test
Thermo Scientific
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
39
Running a Water Test
When a method has been loaded, the following screen will appear.
Press the
Run Test
function key. The following screen appears.
Open the sample compartment. If the desired sample holder is not in the beam, press the
button on the carousel control of the keypad corresponding to the correct cell holder shape
(circle for vials or test-tubes, square for 10 mm square cuvettes, rectangle for long path
rectangular cuvettes). If necessary, the instrument will rotate the carousel to place the correct
cell holder in the measurement position.
Insert a cuvette, tube or vial containing the blank or zero solution into the sample holder.
Note
Ideally, the same cuvette or vial, or one that has been matched to the sample vial,
should be used.
for use of reagent blanks in reverse color methods.