23
Maintenance
Replacing the Tungsten Lamp
124
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
To replace the tungsten lamp
1. Turn off and unplug the instrument from the wall outlet or power strip.
2. Make sure the lamp has cooled (for at least 10 minutes).
3. Loosen the captive screw on the lamp door by rotating it counterclockwise about ¼ turn.
4. Raise the lamp door.
5. Unscrew the thumbscrew while holding the lamp assembly in place.
6. Carefully pull the assembly out until you can reach the connector.
7. Release the tab and disconnect the cable.
8. Unpack the new lamp assembly.
9. Use a soft, lint-free cloth to wipe off any fingerprints on the lamp.
10. Connect the cable to the connector.
WARNING
Turn off and unplug the instrument before removing the cover of the lamp
compartment.
Captive screw on lamp door
Thumbscrew
Lamp assembly
Tab on cable connector
Aligning the lamp assembly