22
Performance Verification
Photometric Accuracy
114
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
Selecting the Mode
To switch between absorbance and %Transmittance, press
Change to %T
(or
Change to
ABS
) until the appropriate mode appears.
Adding Standards
You will need to set three values whenever you add a standard: the wavelength, the absorbance
(or %Transmittance) and the tolerance value.
To add a standard
1. Press
Add Std
and enter the wavelength value in the
Entry
field.
2. Press
Enter
or
Add nm
to add the wavelength to the list.
3. Enter the absorbance or %T value for the entered wavelength in the
Entry
field.
4. Press
Enter
.
5. Enter the tolerance for the entered wavelength in the
Entry
field.
6. Press
Enter
.
The test screen displays the values you just entered for that standard.
7. Press
Start Test
to begin the measurement or
Press Esc
to save the test.
Deleting Standards
To delete a standard
1. Highlight the appropriate standard.