14
Concentration Measurements—Advanced A-%T-C
Entering a Factor
Thermo Scientific
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
63
4. Insert the blank and standard.
5. Press
Enter
to measure the blank and samples and display the absorbance and calculated
factor.
To measure a standard manually (using Manual 6 or Single Cell Holder)
1. Press
Run Standard
.
2. Enter the concentration of the standard and press
Enter
.
3. Insert the blank and standard.
4. Press
Measure Blank
.
If a 6-Position Cell Holder is installed, it moves to the B position to measure the blank
and returns to its previous position.
5. Press
Measure Standard
.
The instrument measures the absorbance of the standard and displays the absorbance and
calculated factor.
Entering a Factor
To enter a factor
Highlight
Factor
. To change the factor, enter the correct factor.
To change the units, highlight
Units
and select the correct units.
Measuring Samples
To measure a sample automatically (using Auto 6 or Auto 3)
1. Press
Run Test
.
2. When prompted, place the blank and samples in their cell positions and press
Enter
.
The instrument measures the blank and samples and displays the sample measurements.
3. Press
Measure Sample
to measure additional samples.
To measure samples manually (using Manual 6 or Single Cell Holder)
1. Press
Run Test
.
2. Insert the blank and sample.
A 6-Position Cell Holder can hold five samples.