21
Kinetics
Recalling and Recalculating Graphical Kinetics Results
102
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
3. Press
Measure Sample
.
After the measurement, the kinetics data and rate appear.
4. If a Single Cell Holder is installed, press
Measure Blank
to measure the blank, insert your
sample and then press
Measure Sample
.
After the measurement, the kinetics data and rate appear.
To switch between tabular and graphical displays, press
Graph/Tabular
.
On a graphical display you can press
Edit Graph
and then press
Cursor
to move the
cursor line from one position to another on the plot. As the cursor moves, the rate and
result values indicate the values for the point where the cursor is located.
Recalling and Recalculating Graphical Kinetics Results
The Kinetics application lets you view and manipulate results in graphical or tabular form.
When your results are displayed, you can modify the range (start and stop time) and the
instrument recalculates the reaction rate.
When working with graphical kinetics results, you need to press
Edit Graph
before you can
rescale and recalculate.