5
Cell Correction
Cell Correction
22
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
3. Highlight
Cell Correction
and press
Enter
.
Cell Correction is now activated, as indicated by the word
On
.
4. Highlight
Setup Correction
and press
Enter
.
5. Highlight
Correction Mode
and press
Enter
to set the mode to either:
Scan – Cell Correction is run on a blank and one sample cell for the range of wavelengths
you specify in Scanning mode.
Discrete nms – Cell Correction is run on a blank and up to five sample cells for up to 31
user-specified, discrete wavelengths.
6. If you selected Scan mode in the preceding step, specify the
Start Wavelength
and the
Stop Wavelength
values.
Note
When Cell Correction is activated, additional parameter lines are added to the
screen above the Cell Correction line. If the Cell Correction line is no longer visible,
highlight
More Parameters
and press
Enter
.