21
Kinetics
Rescaling and Recalculating Tabular Kinetics Results
Thermo Scientific
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
105
The instrument redraws the plot using the selected minimum and maximum values and
displays the recalculated rate and result.
Rescaling and Recalculating Tabular Kinetics Results
When working with tabular kinetics results, you must press
Edit Data
before you can rescale
and recalculate.
After collecting kinetics data, you can use all the data for the rate calculation or select specific
start and end times. When you modify the start and end times or select all the data, the
instrument recalculates and displays the new reaction rate and result.
The edit screen lets you:
• Use all the data to recalculate the results
• Select specific start and end times for the rate calculation and recalculate the results
To use all the data to calculate the reaction rate
With your table of kinetics data displayed on the edit screen, press
Use All Data
.
The instrument calculates and displays the rate.
To select specific start and end times for the rate calculation
1. With your kinetics data displayed on the edit screen, highlight the appropriate data point
in the table.
2. Press
Start Time
or
End Time
to display the recalculated rate and result.