24
Parameters
132
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
Number of Standards
Enters the number of standards to be measured for the
standard curve
Printer
Selects the output mode (RS-232, Parallel)
(Utility)
Protein Factor
Enters the factor to calculate Protein concentration
(DNA Bio Tests)
Ref. Wavelength
Enters a reference wavelength value; for each reported
measurement, measures the analytical wavelength and
reference wavelength. Reported measurement =
Abs@Analytical WL - Abs@Reference WL
Ref. Wavelength
Correction
Turns reference wavelength correction on or off
Run Corr.
Initiates the collection of cuvette data for cell correction
Run Standard
Goes to the Standards entry screen
Run Test
Goes to the data collection screen
(all tests)
Sample Positioner
Selects the type of Positioner
1 Cell
= no movement (zeros and measures sample in same
position
(
Not
available in Kinetics and Scanning)
Manual 6
= cell changer moved by buttons (always zeros on
position B, then returns to set position to start measurement
(
Not
available in Kinetics and Scanning)
Auto 3
= turret auto moved - B, 2, 4 (always zeros on position
B, then goes to position 2 to start measurement)
(
Not
available in Kinetics and Scanning)
Auto 6
= turret auto moved - B,1,2,3,4,5 (always zeros on
position B, then goes to position 1 to start measurement)
Sample Volume
Enters the total volume of sample
(under Dilution Multiplier in some Bio Tests)
Save Test
Saves all parameters of the current test in internal memory for
later recall
(all tests)
Scan speed
Selects the speed (nm/min) for a scan – Slow, Medium, Fast
(Scanning tests only)
Screen Contrast
Lets you improve visibility of the display by changing the
contrast between the background and text
(Utility)
Parameter
Description