Concentration Measurements—Standard Curve Application
Measuring the Standards for a Standard Curve
Thermo Scientific
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
93
3. Set
Entry concentration
and press
Enter
.
4. Press
Measure Standards
.
shows the absorbance of each standard, along with the slope,
intercept and correlation coefficient of the standard curve.
To measure standards manually (using Manual 6 or Single Cell Holder)
1. Install the blank and standards.
2. When the parameters are correct, press
Run Standards
.
3. Set
Entry concentration
.
4. Press
Measure Blank
.
If a 6-Position Cell Holder is installed, it moves to the B position to measure the blank
and returns to its previous position.
5. Press
Measure Standards
.