Operation
Instrument Controls Menu
3-56
Model 410
i
Instruction Manual
Thermo Fisher Scientific
necessary for linear inputs, however a larger number of points may be used
to approximate non-linear inputs. The points range from 2 to 10, with a
default of 2.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel >
Table Points
.
NUMBER OF TABLE POINTS:
CURRENTLY: 2
SET TO: 10 ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Table Point submenu allows the user to set up an individual table
point.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel >
Point 1-10
.
TABLE POINT 01 CONFIG:
>VOLTS 0.00
USER VALUE 0.00
RANGE AVG DIAGS ALARM
The Volts screen allows the user to set the input voltage for the selected
table point in the conversion table, from 0.00 to 10.50. The default table is
a two-point table with point 1: 0.00 V = 000.0 U and point 2: 10.00 V =
10.0 U, where U is the previously entered unit of measure.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel > select Point >
Volts
.
TABLE POINT 01 VOLTS:
CURRENTLY: 0.00
SET TO: 00.0
0
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Table Point
Volts