Operation
Diagnostics Menu
3-66
Model 410
i
Instruction Manual
Thermo Fisher Scientific
ANALOG INPUT READINGS:
>NO 10.2 ppm
NO2 18.2 ppb
CO2 250 ppm
FL1 20.42 L/min
IO7 0.0 v
RANGE AVG DIAGS ALARM
The Analog Input Voltages screen (read only) displays the 8 raw analog
voltage readings (if the I/O expansion board option is installed).
●
In the Main Menu, choose Diagnostics >
Analog Input Voltages
.
ANALOG INPUT VOLTAGES:
>ANALOG IN 1 6.24 V
ANALOG IN 2 4.28 V
ANALOG IN 3 0.00 V
ANALOG IN 4 0.00 V
ANALOG IN 5 0.00 V
RANGE AVG DIAGS ALARM
The Digital Inputs screen (read only) displays the state of the 16 digital
inputs. Pull-ups are provided on all of the inputs, so if nothing is connected
they will read (1), if an input is brought to ground, it will read (0).
●
In the Main Menu, choose Diagnostics >
Digital Inputs
.
DIGITAL INPUTS:
>INPUT 1 1
INPUT 2 1
INPUT 3 1
INPUT 4 1
INPUT 5 1
RANGE AVG DIAGS ALARM
The Relay States screen displays the state of the 10 digital outputs and
allows toggling of the state to either on (1) or off (0). The relays are
restored to their original states upon exiting this screen.
●
In the Main Menu, choose Diagnostics >
Relay States
.
●
Press
to toggle and set the relay state to open or closed for the
selected digital output.
Analog Input
Voltages
Digital Inputs
Relay States