C-Link Protocol Commands
Calibration
Thermo Fisher Scientific
Model 410
i
Instruction Manual
B-33
zero dur
This command reports the zero duration. The following example reports
that the zero duration is 10 minutes.
Send:
zero dur 10
Receive:
zero dur 10 ok
set zero dur
value
This command sets the zero duration to
value
where
value
represents the
zero duration in minutes. The following example sets the zero duration to
15 minutes.
Send:
set zero dur 15
Receive:
set zero dur 15 ok
zs period
This command reports the zero/span (zs) period. The following example
reports that the zero/span period is 24 hours.
Send:
zs
period
Receive:
zs period 24 hr
set zs period
value
This command sets zero/span (zs) period to
value
, where
value
represents
the zero/span period in hours. The following example sets the zero/span
period to 24 hours.
Send:
set zs period 24
Receive:
set zs period 24 ok
zs avg time
This command reports the zero/span (zs) averaging time in seconds. The
following example reports that the zero/span averaging time is 30 seconds,
according to
Table B–3
.
Send:
zs
avg
time
Receive:
zs avg time 5:30 sec
set avg time
value
This command sets the zero/span averaging time, according to
Table B–3
.
The following example sets the zero/span averaging time to 120 seconds.
Send:
set zs avg time 8
Receive:
set zs avg time 8 ok