Operation
Alarms Menu
Thermo Fisher Scientific
Model 410
i
Instruction Manual
3-71
the minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Pressure
.
PRESSURE:
ACTUAL 753.9 mmHg
>MIN 250.0 mmHg
MAX 1000.0 mmHg
RANGE AVG DIAGS ALARM
The Minimum Pressure alarm limit screen is used to change the minimum
pressure alarm limit. The minimum and maximum pressure screens
function the same way.
●
In the Main Menu, choose Alarms > Pressure >
Min
or
Max
.
PRESSURE:
ACTUAL MIN: 250.0 mmHg
SET MIN TO: 500.0 mmHg ?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Sample Flow screen displays the current sample flow reading and sets
the minimum and maximum alarm limits. Acceptable alarm limits range
from .350 to 1.500 L/min. If the sample flow reading goes beyond either
the minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Sample Flow
.
SAMPLE FLOW:
ACTUAL 1.108 L/min
>MIN 0.350 L/min
MAX 1.500 L/min
RANGE AVG DIAGS ALARM
Min and Max Pressure
Limits
Sample Flow