C-Link Protocol Commands
I/O Configuration
Thermo Fisher Scientific
Model 410
i
Instruction Manual
B-53
list din
list dout
These commands report the current selection for the digital inputs or the
digital outputs in the format. Output no Index number variable name
active state. The active state for digital outputs is open or closed. The active
state for digital inputs is high or low. Channels that have not been assigned
a variable index are not displayed.
Send:
list
dout
Receive: list
dout
output
index
variable
state
1 22 I/O BD COMM closed
2
2
LOCAL/REMOTE
open
3
4
UNITS
open
4 19 BIAS VOLTAGE closed
7
7
SAMPLE
MODE
open
8 8 GEN MODE open
list var aout
list var dout
list var din
These commands report the list of index numbers, and the variables
(associated with that index number) available for selection in the current
mode for analog outputs, digital outputs, and digital inputs. The index
number is used to insert the variable in a field location in a list using “set sp
field index
”. The following example reports the list of analog outputs, index
numbers, and variables.
Send:
list var aout
Receive:
list var aout
index
variable
0
none
5
sr
8
intt
9
cht
10
auxt
13
pres
14
smplfl
15
intensity
16
motor
18
ain1
19
ain2
20
ain3
21
ain4
22
ain5
23
ain6
24
ain7
25
co2
Содержание 410i
Страница 1: ...Model 410i Instruction Manual Optical Filter CO2 Analyzer Part Number 102011 00 25Jul2017 ...
Страница 4: ......
Страница 35: ...Installation Connecting External Devices Thermo Fisher Scientific Model 410i Instruction Manual 2 9 ...
Страница 142: ......
Страница 216: ......
Страница 300: ......
Страница 321: ......