Operation
Instrument Controls Menu
3-38
Model 410
i
Instruction Manual
Thermo Fisher Scientific
INSTRUMENT ID:
CURRENTLY: 68
SET TO: 70 ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
A Gesytec serial number is included in the response to a Gesytec protocol
DA command. Refer to “
Data Sampling/Data Query Command (DA)
” on
page
D-4
for detailed information and examples.
A Gesytec serial number defaults to zero.
●
In the Main Menu, choose Instrument Controls > Communication
Settings >
Gesytec Serial No
.
GESYTEC SERIAL NO:
CURRENTLY: 0
SET TO: 0 ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Communication Protocol screen is used to change the protocol for
serial communications. Possible settings are C-Link, MODBUS, Gesytec,
and Streaming.
C-link Protocol
- C-link is a bi-directional protocol that provides access to
all instrument functions including reading the various concentrations and
other analog values or variables, reading the status of the digital outputs of
the instrument, and triggering or simulating the activation of a digital
input to the instrument. Refer to Appendix B for detailed C-link
information.
Streaming Protocol
- The streaming protocol provides a one-way reporting
capability in which data records are exported to a serial device on a regular
basis without prompting.
Gesytec (Bayern-Hessen) protocol
- The Gesytec (Bayern-Hessen)
protocol support for the Model 410
i
enables the user to read the various
concentrations. Refer to Appendix D for detailed Gesytec (Bayern-Hessen)
information.
Gesytec Serial No
Communication Protocol