C-Link Protocol Commands
Datalogging
Thermo Fisher Scientific
Model 410
i
Instruction Manual
B-23
10:15 05-12-03 flags 9c040000 CO2 7349E+0 loCO2 5994E+0
intt 33.2 cht 44.7 pres 758.9 smplfl 1.085 speed 100.0
biasv -115.5 intensity 1999940
10:15 05-12-03 flags 9c040000 CO2 7349E+0 loCO2 5994E+0
intt 33.2 cht 44.7 pres 758.9 smplfl 1.085 speed 100.0
biasv -115.5 intensity 1999940
10:15 05-12-03 flags 9c040000 CO2 7349E+0 loCO2 5994E+0
intt 33.2 cht 44.7 pres 758.9 smplfl 1.085 speed 100.0
biasv -115.5 intensity 1999940
lrec format
srec format
erec format
These commands report the output format for lrecs and srecs, and erec data
in various formats such as ASCII without text, ASCII with text, or binary.
The following example shows the output format for lrecs is ASCII with
text, according to
Table B–5
.
Send:
lrec
format
Receive: lrec
format
1
set lrec format
format
set srec format
format
set erec format
format
These commands set the output format for lrecs and srecs, and erec data,
according to
Table B–5
. The following example sets the lrec output format
to ASCII with text.
Send:
set lrec format 1
Receive:
set lrec format 1 ok
Table B–5.
Record Output Formats
Format
Output Format
0
ASCII no text
1
ASCII with text
2 Binary
data
lrec layout
srec layout
erec layout
These commands report the layout (string indicating the data formats) for
data that is sent out in response to the erec, lrec, srec, and related
commands. For details on how to interpret the strings, see “Record Layout
Definition” later in this appendix.