Operation
Service Menu
Thermo Fisher Scientific
Model 410
i
Instruction Manual
3-79
RANGE MODE SELECT:
CURRENTLY: SINGLE
SET TO: DUAL ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Pressure Calibration submenu is used to calibrate the pressure sensor
to zero, span, or restore factory default values. The pressure calibration is
visible only when the instrument is in service mode. For more information
on the service mode, see “
Service Mode
” earlier in this chapter.
The pressure sensor’s zero counts and span slope are displayed on the
menu.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Pressure Calibration
.
PRESSURE SENSOR CAL:
>ZERO 72
SPAN 1.1416
SET DEFAULTS
RANGE AVG DIAGS ALARM
The Calibrate Pressure Zero screen calibrates the pressure sensor at zero
pressure.
Note
A vacuum pump must be connected to the pressure sensor before
performing the zero calibration.
▲
Note
Wait at least 30 seconds for the reading to stabilize before saving the
value.
▲
●
In the Main Menu, choose Service > Pressure Calibration >
Zero
.
Pressure Calibration
Calibrate Pressure Zero
Содержание 410i
Страница 1: ...Model 410i Instruction Manual Optical Filter CO2 Analyzer Part Number 102011 00 25Jul2017 ...
Страница 4: ......
Страница 35: ...Installation Connecting External Devices Thermo Fisher Scientific Model 410i Instruction Manual 2 9 ...
Страница 142: ......
Страница 216: ......
Страница 300: ......
Страница 321: ......