Servicing
Pressure Transducer Replacement
7-24
Model 410
i
Instruction Manual
Thermo Fisher Scientific
3.
Remove both lines from the pump.
4.
Loosen the four captive screws holding the pump bracket to the shock
mounts and remove the pump assembly and the shock mounts.
5.
Invert the pump assembly, remove the two pump mounting screws
located on the bottom side of the pump bracket, and remove the
bracket from the pump.
6.
Install the new pump by following the previous steps in reverse.
7.
Perform a leak test as described in the “
Preventive Maintenance
”
chapter.
Figure 7–15.
Replacing the Pump
Use the following procedure to replace the pressure transducer (
Figure 7–
16
).
Equipment Required:
Pressure transducer
Philips screwdriver, #2
Pressure Transducer
Replacement
Pump
Pump Bracket
Pump Mounting Screws (2)
(Bottom)
Captive Screws (4)
Shock Mounts (4))
Floor Plate