Gesytec (Bayern-Hessen) Protocol
Gesytec Commands
D-8
Model 410
i
Instruction Manual
Thermo Fisher Scientific
Table D–2.
Error Status for Model 410
i
D7 D6 D5 D4 D3 D2 D1 D0
Bit
8
7
6
5
4
3
2
1
Hex-value
80 40 20 10 08 04 02 01
MSB
LSB
Error status:
Internal
Temperature
Alarm
0 0 0 0 0 0 0 1
Optical Chamber Temperature
Alarm
0 0 0 0 0 0 1 0
AGC
Intensity
Alarm
0 0 0 0 0 1 0 0
Bias
Voltage
Alarm
0 0 0 0 1 0 0 0
Pressure
Alarm
0 0 0 1 0 0 0 0
Sample
Flow
Alarm
0 0 1 0 0 0 0 0
Motor
Speed
Alarm
0 1 0 0 0 0 0 0
Not
Used
1 0 0 0 0 0 0 0