C-Link Protocol Commands
Calibration
B-28
Model 410
i
Instruction Manual
Thermo Fisher Scientific
set cal co2 coef
set cal high co2 coef
set cal low co2 coef
set span calibrate
set calibrate
These commands will set CO
2
coefficients based on CO
2
span gas
concentrations. The high and low commands are only available in dual and
auto range mode. If the mode is incorrect, the instrument responds with
“can’t, wrong settings”. The following example shows a successful auto-
calibration of the low CO
2
coefficient.
Note
Span gas must be provided to the instrument prior to issuing any of
these commands.
▲
Send:
set cal low CO2 coef
Receive:
set cal low CO2 coef ok
cal coef o2
This command will set O
2
coefficient based on O
2
span gas concentrations.
The example below shows a successful auto-calibration of the O
2
coefficient.
Send:
set cal coef o2
Receive:
set cal coef o2 ok
co2 coef
high co2 coef
low co2 coef
These commands report the current CO
2
coefficient in single range mode,
or the high and low range coefficient in dual or auto range mode. If the
mode is incorrect, the instrument responds with “can’t, wrong settings”.
The following example reports that the CO
2
coefficient is 1.200.
Send:
CO2
coef
Receive: CO2
coef
1.200
set co2 coef
value
set high co2 coef
value
set low co2 coef
value
These commands set the CO
2
coefficients to a user-defined
value
, where
value
is a floating-point representation of the coefficient. The following
example sets the CO
2
coefficient to 1.200.
Send:
set CO2 coef 1.200
Receive:
set CO2 coef 1.200 ok