Installation
Startup
2-10
Model 410
i
Instruction Manual
Thermo Fisher Scientific
Use the following procedure when starting the instrument.
1.
Turn the power ON.
2.
Set the PURGE port input pressure to 20 psig.
3.
Allow 90 minutes for the instrument to stabilize.
4.
Set instrument parameters such as operating ranges and averaging times
to appropriate settings. For more information about instrument
parameters, see the “
Operation
” chapter.
5.
Before beginning the actual monitoring, perform a multipoint
calibration as described in the “
Calibration
” chapter.
Startup