The Configure Datalogging menu deals with datalogging configuration for
the currently selected record type.
●
In the Main Menu, choose Instrument Controls > Datalogging Settings
>
Configure Datalogging
.
DATALOGGING SETTINGS:
>LOGGING PERIOD MIN 60
MEMORY ALLOCATION % 50
DATA TREATMENT AVG
FLAG STATUS DATA OFF
RANGE AVG DIAGS ALARM
The Logging Period Min screen is used to select the logging period in
minutes for the selected record format (srec or lrec). List of choices include:
off, 1, 5, 15, 30, and 60 minutes (default).
●
In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging >
Logging Period Min
.
SET PERIOD FOR SREC:
CURRENTLY: 60 MIN
SET TO: OFF MIN ?
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Memory Allocation Percent screen is used to select the percentage of
total memory that may be used by the selected record type (lrecs or srecs).
Percentages between 0 and 100% are available in increments of 10.
Changing this value results in log erasure for both types of records, and
changing the percent allocated to one record type will automatically change
the other.
●
In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging >
Memory Allocation %
.
Configure Datalogging
Logging Period Min
Memory Allocation
Percent