Operation
Diagnostics Menu
Thermo Fisher Scientific
Model 410
i
Instruction Manual
3-67
RELAY STATE:
>OUTPUT 1 0
OUTPUT 2 0
OUTPUT 3 0
OUTPUT 4 1
OUTPUT 5 1
RANGE AVG DIAGS ALARM
The Test Analog Outputs menu is used to set each of the analog output
channels to zero or full-scale. Channel choices include all analog outputs, 6
voltage channels, and 6 current channels (if the I/O expansion board
option is installed).
●
In the Main Menu, choose Diagnostics >
Test Analog Outputs
.
TEST ANALOG OUTPUTS:
>ALL
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
RANGE AVG DIAGS ALARM
The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-
scale voltage, zero sets the analog outputs to 0 volts, and reset returns the
output to normal operation. The analog outputs are returned to normal
operation upon exiting this screen. The following example shows the screen
when all analog outputs are set to “normal” operating mode.
●
In the Main Menu, choose Diagnostics > Test Analog Outputs >
ALL,
Voltage Channel 1–6
,
or
Current Channel 1–6
.
SET ANALOG OUTPUTS:
SETTING: ALL
OUTPUT SET TO: NORMAL
SET TO FULL SCALE
SET TO ZERO
RESET TO NORMAL
RANGE AVG DIAGS ALARM
The Instrument Configuration screen displays information on the
hardware configuration of the instrument.
Test Analog Outputs
Set Analog Outputs
Instrument
Configuration