Servicing
Optical Bench Replacement
Thermo Fisher Scientific
Model 410
i
Instruction Manual
7-17
Optical Bench
Shock Mounts (6)
Optical Bench
Shock Mounts (6)
4.
Remove the four screws holding the optical bench to the shock mounts
and carefully remove the optical bench.
5.
Replace the optical bench by following the previous steps in reverse.
6.
Calibrate the instrument. Refer to the “
Calibration
” chapter in this
manual.
Figure 7–9.
Replacing the Optical Bench (Standard)
Figure 7–10.
Replacing the Optical Bench (High Level)