MODBUS Protocol
TCP Communication Parameters
C-2
Model 410
i
Instruction Manual
Thermo Fisher Scientific
i
Series instruments support the MODBUS/TCP protocol. The register
definition is the same as for the serial interface. Up to three simultaneous
connections are supported over Ethernet.
TCP connection port for MODBUS : 502
Here are the MODBUS ADU (Application Data Unit) formats over serial
and TCP/IP:
Serial:
Slave Address
Function Code
Data
Error Check
TCP/IP: MBAP Header Function Code
Data
The MODBUS slave address is a single byte in length. This is the same as
the instrument ID used for C-Link commands and can be between 1 and
127 decimal (i.e. 0x01 hex to 0x7F hex). This address is only used for
MODBUS RTU over serial connections.
Note
Device ID ‘0’, used for broadcast MODBUS commands, is not
supported. Device IDs 128 through 247 (i.e. 0x80 hex to 0xF7 hex) are not
supported because of limitations imposed by C-Link.
▲
In MODBUS over TCP/IP, a MODBUS Application Protocol Header
(MBAP) is used to identify the message. This header consists of the
following components:
Transaction Identifier
2 Bytes
0x0000 to 0xFFFF (Passed back in response)
Protocol Identifier
2 Bytes
0x00 (MODBUS protocol)
Length
2 Bytes
0x0000 to 0xFFFF (Number of following bytes)
Unit Identifier
1 Byte
0x00 to 0xFF (Passed back in response)
TCP Communication
Parameters
Application Data
Unit Definition
Slave Address
MBAP Header
Содержание 410i
Страница 1: ...Model 410i Instruction Manual Optical Filter CO2 Analyzer Part Number 102011 00 25Jul2017 ...
Страница 4: ......
Страница 35: ...Installation Connecting External Devices Thermo Fisher Scientific Model 410i Instruction Manual 2 9 ...
Страница 142: ......
Страница 216: ......
Страница 300: ......
Страница 321: ......