Operation
Alarms Menu
Thermo Fisher Scientific
Model 410
i
Instruction Manual
3-75
●
In the Main Menu, choose Alarms >
Zero
or
Span Autocal
.
ZERO AUTO CALIBRATION:
ALARM: OK
RESPONSE: 5.0
RANGE AVG DIAGS ALARM
The Concentration screen displays the current CO
2
concentration and sets
the minimum and maximum alarm limits. In the standard model,
acceptable alarm limits range from 0 to 10000 ppm. In the high level
model, acceptable alarm limits range from 0 to 250000 ppm, or 1 to 25%.
The minimum alarm may be programmed as a floor trigger (alarm is
triggered when the concentration falls below the minimum value) or a
ceiling trigger (alarm is triggered when the concentration goes above the
minimum value). If the CO
2
concentration goes beyond either the
minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms >
Concentration
.
CONCENTRATION:
ACTUAL 15.8
>MIN 0.0
MAX 10000
MIN TRIGGER CEILING
RANGE AVG DIAGS ALARM
The Minimum Concentration alarm limit screen is used to change the
minimum concentration alarm limits. The minimum and maximum
concentration alarm limit screens function the same way.
●
In the Main Menu, choose Alarms > Concentration >
Min
or
Max
.
CONCENTRATION:
ACTUAL MIN: 0.0
SET MIN TO: 0002
0
.00 ?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Concentration
Min and Max
Concentration Limits