Operation
Alarms Menu
3-74
Model 410
i
Instruction Manual
Thermo Fisher Scientific
MOTOR SPEED:
ACTUAL 100.0 %
RANGE AVG DIAGS ALARM
The Zero and Span Check screens allow the user to view the status of the
most recent zero/span checks and set the maximum check offsets. An alarm
will be triggered if a zero or span check indicates drift that exceeds the
offset value. The zero and span check screens are visible only if the
zero/span check option is enabled. Their functions are similar.
●
In the Main Menu, choose Alarms >
Zero
or
Span Check
.
ZERO CHECK:
ALARM: OK
RESPONSE: 0.0
>MAX OFFSET 10.0
RANGE AVG DIAGS ALARM
The Max Zero Check Offset screen is used to change the maximum zero
check offset. The maximum zero and span offset screens function the same
way.
●
In the Main Menu, choose Alarms > Zero or Span Check >
Max
Offset
.
MAX ZERO CHECK OFFSET:
CURRENTLY: 10.0
SET TO: 0001
1
.00 ?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Zero and Span Auto Calibration screens (read only) allow the user to
view the status of the most recent auto background calibration or span
calibrations. The zero and span auto calibration screens are visible only if
the zero/span check option is enabled and the zero or span cal reset
function is enabled and the zero and span cal reset function is enabled.
Zero and Span
Check
Max Zero and Span Offset
Zero and Span Auto
Calibration