C-Link Protocol Commands
Calibration
B-30
Model 410
i
Instruction Manual
Thermo Fisher Scientific
Send:
set co2 bkg 1.400
Receive:
set co2 bkg 1.400 ok
bkg o2
This command reports the oxygen background concentration in percent.
The following example reports that the current O
2
background
concentration is 1.5%.
Send:
bkg
o2
Receive:
bkg o2 1.50 %
set bkg o2
This command is used to set the oxygen background concentration in
percent. The example below sets the O
2
background concentration to
1.5%.
Send:
set bkg o2 1.5
Receive:
set bkg o2 1.5 ok
init ratio
This command reports the initial sample/reference ratio. The following
example reports that the initial ratio was 1.16210.
Send:
initial
ratio
Receive:
initial ratio 1.16210
sp conc
high sp conc
low sp conc
These commands report span concentration in single range mode, or the
high and low span concentrations in dual or auto range mode. If the mode
is incorrect, the instrument responds with “can’t, wrong settings”. The
following example reports the span gas concentration in single range mode.
Send:
sp
conc
Receive:
sp conc 1000
set sp conc
value
set high sp conc
value
set low sp conc
value
These commands set the span concentration to a user-defined
value
, where
value
is a floating-point representation of the span concentration in the
currently selected units. The following example sets the span concentration
to 1000 ppm in the single range mode.
Send:
set sp conc 1000
Receive:
set sp conc 1000 ok