41
3.1.6 Range (Manual & Auto)
Definition
When making measurements except CONT, DIODE and Temperature, you can let the
machine choose ranges for you, or you can select the appropriate ranges manually by
yourself. The difference between auto-range and manual-range is the settling time.
Auto-range is a convenient way for user, but manual-range can usually speed up the
process.
If the input signal is beyond allowed range, an “OVLD” message will be shown on the
display. The threshold of maximum/minimum readings for each range are 120 % of the
range for maximum and 10% of the range for minimum.
Default
The default is auto-range. The user selected range is stored in volatile memory and the
default will be restored when the meter is power-off. Please refer to
Table 2-1
on
page
14 for the factory default range
.
How to set the auto / manual range:
You can set the auto / manual range either through the front panel operation or through
the remote interface operation.
Front Panel Operation
First choose a measurement function on the front panel, then press “AUTO” button to
select auto-range feature. Or use
△
and
▽
buttons to manually select a range. If
selected range is not shown on the bottom of the display, press round PREV or NEXT
buttons until the range information appears. The locations of these buttons are shown
with a red rectangle frame in
Figure 4-6
.
Figure 4-6
Remote Interface Operation
Use the following commands from your PC terminal to set the range:
Содержание DL-2060
Страница 1: ...B71 0110 41 INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 2060 DL 2060G DL 2060VG DL 2060R DL 2060VR...
Страница 37: ...31 Figure 3 17 1 2 3 5 5 4...
Страница 135: ...129 C Tree Diagram Please refer to G Notice of Scanner Card if your firmware version is 1 05 or upward...
Страница 136: ...130...
Страница 137: ...131...
Страница 138: ...132 D DL 2060 s Dimension...
Страница 152: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...