122
TRIGger:SOURce?
Query the trigger source.
TRIGger:DELay {<
seconds
>|MINimum|MAXimum}
Set a trigger delay time in seconds. The delay is the time between the trigger signal and each
sample that follows. Specify a delay time from 0 to 3600 seconds.
TRIGger:DELay?
Query the trigger delay time.
TRIGger:DELay:AUTO {OFF|ON}
Disable or enable a automatic trigger delay. The delay is determined by function, range,
integration time, and ac filter setting. Specifying a delay time automatically turns off the
automatic trigger delay.
TRIGger:DELay:AUTO?
Query the automatic trigger delay mode. Returns “0” (OFF) or “1” (ON).
SAMPle:COUNt {<
value
>|MINimum|MAXimum}
Set the number of readings (samples) the multimeter takes per trigger. Select a number from 1
to 50,000 readings per trigger.
SAMPle:COUNt ? [MINimum|MAXimum]
Query the sample count.
TRIGger:COUNt {<
value
>|MINimum|MAXimum|INFinite}
Specify the number of triggers the multimeter will accept before returning to the “idle” state.
Select from 1 to 50,000 triggers. The INFinite parameter instructs the multimeter to take triggers
continuously. Trigger count is ignored while in local operation.
TRIGger:COUNt? [MINimum|MAXimum|INFinite]
Query the trigger count. Returns “9.9037” when you select the infinite trigger count.
Содержание DL-2060
Страница 1: ...B71 0110 41 INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 2060 DL 2060G DL 2060VG DL 2060R DL 2060VR...
Страница 37: ...31 Figure 3 17 1 2 3 5 5 4...
Страница 135: ...129 C Tree Diagram Please refer to G Notice of Scanner Card if your firmware version is 1 05 or upward...
Страница 136: ...130...
Страница 137: ...131...
Страница 138: ...132 D DL 2060 s Dimension...
Страница 152: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...