The moving average filter puts the specified number of reading conversions in a first-in,
first-out order. The very first measurement reading simply fills up the stack. To yield a
reading for display, the filter produces an average of the stacked measurement readings
every time a new measurement reading is available and replaces the oldest reading in the
stack. In the repeating average mode, the multimeter waits for the measurement reading
stack to fill up and then take an average to produce a reading for display. It then flushes
the stack and starts over with an empty stack. Consequently, the repeating digital filter
yields one reading for display every specified number of measurement readings.
Digital filter is not available for diode, continuity, frequency and period measurements.
Default
The digital filter is enabled and is in moving average mode with 10 readings by default.
How to enable/disable digital filter
Press FILTER button to switch the digital filter function. The “FILT” anunnciator
indicates the state of the digital filter. When it is lit, the filter is enabled.
How to configure digital filter:
You can configure the digital Filter either through the front panel operation or through
the remote interface operation.
Front Panel Operation
For READINGS setting, please press CONFIG and then FILTER button for selecting
READINGS. Press ENTER to access. Use
◁
,
▷
,
△
and
▽
buttons to set the READINGS
(from 2 to 100) in order to get an average value. Press ENTER when it’s ready.
For MODE setting, please press CONFIG and then FILTER button for selecting MODE. Press
ENTER to access. Use
◁
,
▷
to select MOVING AVG. or REPEAT AVG.. Please ENTER to
confirm.
Press ESC button to enable or disable this function.
Remote Interface Operation
Use the following commands to set your digital filter:
SENSe:AVERage:TCONtrol {MOVing|REPeat}
SENSe:AVERage:TCONtrol?
SENSe:AVERage:COUNt {<
value
>|MINimum|MAXimum}
SENSe:AVERage:COUNt? [MINimum|MAXimum]
SENSe:AVERage:STATe {OFF|ON}
SENSe:AVERage:STATe?
Содержание DL-2060
Страница 1: ...B71 0110 41 INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 2060 DL 2060G DL 2060VG DL 2060R DL 2060VR...
Страница 37: ...31 Figure 3 17 1 2 3 5 5 4...
Страница 135: ...129 C Tree Diagram Please refer to G Notice of Scanner Card if your firmware version is 1 05 or upward...
Страница 136: ...130...
Страница 137: ...131...
Страница 138: ...132 D DL 2060 s Dimension...
Страница 152: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...