66
Figure 4-23
How to make a
dBm measurement
Select a measurement function by pressing DCV or ACV button. The dBm feature is
only available for DCV and ACV only.
To enable dBm, press SHIFT + dBm(MIN/MAX) buttons.
After enabling the dBm operation, user can select the reference resistance as
mentioned above. The default for the reference resistance is 600 Ω.
The locations of the buttons are shown with red rectangle frames in
Figure 4-24
.
※
Note: Press SHIFT+dBm(MIN/MAX) again to disable this feature. The “MATH”
anunnciator on the display indicates the state of a mathematical feature.
Figure 4-24
Remote Interface Operation
Use the following commands to enable dBm feature or to set the reference resistance:
CALCulate:FUNCtion DBM
CALCulate:STATe {OFF|ON}
CALCulate:STATe?
CALCulate:DBM:REFerence {<
value
>|MINimum|MAXimum}
or