110
MEASure:CURRent:AC? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a AC current measurement with the specified range and resolution. The reading
is sent to the output buffer. For AC measurement, resolution is fixed at 6 ½ digits. Therefore the
resolution parameter only affects the front panel display.
MEASure:RESistance? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a 2-wire Ω measurement with the specified range and resolution. The reading is
sent to the output buffer.
MEASure:FRESistance? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a 4-wire Ω measurement with the specified range and resolution. The reading is
sent to the output buffer.
MEASure:FREQuency? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a frequency measurement with the specified range and resolution. The reading
is sent to the output buffer. For frequency measurements, the meter uses only one “range” for all
inputs between 3Hz and 300kHz. With no input signal applied, frequency measurements return
“0”.
MEASure:PERiod? {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and make a period measurement with the specified range and resolution. The reading is
sent to the output buffer. For period measurements, the meter uses only one “range” for all
inputs between 0.33 seconds and 3.3
μ
sec. With no input signal applied, period measurements
return “0”.
MEASure:CONTinuity?
Preset and make a continuity measurement. The reading is sent to the output buffer. The range
and resolution are fixed at 1kΩ and 5 ½ digits respectively.
MEASure:DIODe?
Preset and make a diode measurement. The reading is sent to the output buffer. The range and
resolution are fixed at 1Vdc with 1mA current source and 5 ½ digits respectively.
MEASure:TEMPerature?
Preset and make a temperature measurement with RTD. The reading is sent to the output buffer.
MEASure:TCOuple?
Preset and make a temperature measurement with thermocouple. The reading is sent to the
output buffer.
Содержание DL-2060
Страница 1: ...B71 0110 41 INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 2060 DL 2060G DL 2060VG DL 2060R DL 2060VR...
Страница 37: ...31 Figure 3 17 1 2 3 5 5 4...
Страница 135: ...129 C Tree Diagram Please refer to G Notice of Scanner Card if your firmware version is 1 05 or upward...
Страница 136: ...130...
Страница 137: ...131...
Страница 138: ...132 D DL 2060 s Dimension...
Страница 152: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...