45
Choosing
SPRTD
takes you to a menu where you can specify the seven coefficients that are
used to determine the temperature. Use
◁
and
▷
to move through the digits and
△
and
▽
to change the numbers to a desired value. Press ENTER to set the value.
Remote Interface Operation
Use the following commands to set up the RTD configuration.
SENSe:TEMPerature:RTD:TYPE{PT100|D100|F100|PT385|PT3916|USER|SPRTD|NTCT}
SENSe:UNIT {Cel|Far|K}
SENSe:UNIT?
SENSe:TEMPerature:RTD:RZERo {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:ALPHa {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:BETA {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:DELTa {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:RZERo {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:A4 {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:B4 {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:AX {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:BX {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:CX {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:SPRTD:DX {<
value
>|MINimum|MAXimum}
Thermocouple
Definition
If users are using the thermocouple function, the selections on the DL-2060 are: type E, J, K,
N, R, S and T. (On the other hand, other selection “SIMULATED” simulates the reference
junction.) To use the the thermocouple function, users have to adjust something. First of all,
via a thermocouple adaptor, users need to measure a known and accurate reference
temperature (T1) as a criterion. Then the display of the DL-2060 will show a reference value
(T2). At the moment, a remainder (T3) from (T1 – T2) will be calculated. After that, users just
adjust the default (simulated value 23.0°C) to 23.0°C + (T3). Under the circumstance, the
thermocouple function will be activated completely.
For Example:
If the criterion value is (T1 =30.0°C), the reference value showing on the DL-2060 is
(T2=24.0°C). Users will get a remainder T3 (T1 – T2 = 6.0°C). Then users just configure the
default simulated value from 23.0°C to 29.0°C on the DL-2060.
How to set up thermocouple
You can set up the thermocouple configuration either through the front panel operation
or through the remote interface operation.
Содержание DL-2060
Страница 1: ...B71 0110 41 INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 2060 DL 2060G DL 2060VG DL 2060R DL 2060VR...
Страница 37: ...31 Figure 3 17 1 2 3 5 5 4...
Страница 135: ...129 C Tree Diagram Please refer to G Notice of Scanner Card if your firmware version is 1 05 or upward...
Страница 136: ...130...
Страница 137: ...131...
Страница 138: ...132 D DL 2060 s Dimension...
Страница 152: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...