71
Figure 4-29
Remote Interface Operation
Users can use the following commands from their PC terminals to store or retrieve
readings in the
memory
.
In addition, the number for STORE function only can be set
through front panel.
INITiate
(
This command tells the meter to be on “wait-for-trigger” state. After a
measurement is taken, measurement readings
will be placed in the memory.
)
FETCh?
(
Use this command to retrieve stored readings.
)
DATA:POINts?
(
Use this command to query the number of stored readings
.)
3.4.4 Sensitivity Band (Hold)
The reading hold function captures and holds a stable reading on the display. The
multimeter beeps and holds the value when it detects a stable reading. The sensitivity
band in reading hold decides which reading is stable enough. This band is expressed as
a percent of reading on the selected range. The multimeter considers a reading stable
when three consecutive values are within the band. User has an option to adjust this
sensitivity band.
Default
The default band is 0.1%. User’s selection is stored in a volatile memory and it will be
cleared after the meter has been turned off.
How to adjust the sensitivity band
You can adjust the sensitivity band through the front panel operation.
Front Panel Operation
In order to adjust the sensitivity band, please press MENU button then use
◁
and
▷
buttons to find “TRIG” submenu. Press ENTER to select it. Then again use
◁
and
▷
buttons to find “READ HOLD” submenu. Press ENTER to choose it. Then use
◁
and
▷
buttons to choose a desired band. Press ENTER to set the desired sensitivity band. The
locations of the buttons are shown with red rectangle frames in
Figure 4-30
.
Содержание DL-2060
Страница 1: ...B71 0110 41 INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 2060 DL 2060G DL 2060VG DL 2060R DL 2060VR...
Страница 37: ...31 Figure 3 17 1 2 3 5 5 4...
Страница 135: ...129 C Tree Diagram Please refer to G Notice of Scanner Card if your firmware version is 1 05 or upward...
Страница 136: ...130...
Страница 137: ...131...
Страница 138: ...132 D DL 2060 s Dimension...
Страница 152: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...