107
※
Note: If you send two query commands without reading the response from the first,
and then attempt to read the second response, you may receive some data from the first
response followed by the complete second response. To avoid this, do not send a query
command without reading the response. When you cannot avoid this situation, send a
device clear before sending the second query command.
SCPI Command Terminators
A command string sent to the multimeter must terminate with a <new line> character.
The GPIB EOI (end-or-identify) message is interpreted as a <new line> character and
can be used to terminate a command string in place of a <new line> character. A
<carriage return> followed by a <new line> is also accepted. Command string
termination will always reset the current SCPI command path to the root level.
GPIB(IEEE-488.2) Common Commands
The GPIB (IEEE-488.2) standard defines a set of common commands that perform
functions like reset, self-test, and status operations. Common commands always begin
with an asterisk ( * ), are four to five characters in length, and may include one or more
parameters. The command keyword is separated from the first parameter by a blank
space.
Use a semicolon ( ; ) to separate multiple commands as shown below:
"*RST; *CLS; *ESE 32; *OPC?"
SCPI Parameter Types
The SCPI language defines several different data formats to be used in program
messages and response messages.
Numeric Parameters
Commands that require numeric parameters will accept all
commonly used decimal representations of numbers including optional signs, decimal
points, and scientific notation.
Special values for numeric parameters like MINimum, MAXimum, and DEFault are also
accepted. You can also send engineering unit suffixes with numeric parameters (e.g., M,
K, or u). If only specific numeric values are accepted, the multimeter will automatically
round the input numeric parameters. The following command uses a numeric
parameter:
VOLTage:DC:RANGe {<range>|MINimum|MAXimum}
Содержание DL-2060
Страница 1: ...B71 0110 41 INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 2060 DL 2060G DL 2060VG DL 2060R DL 2060VR...
Страница 37: ...31 Figure 3 17 1 2 3 5 5 4...
Страница 135: ...129 C Tree Diagram Please refer to G Notice of Scanner Card if your firmware version is 1 05 or upward...
Страница 136: ...130...
Страница 137: ...131...
Страница 138: ...132 D DL 2060 s Dimension...
Страница 152: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...