102
Function
Range
Resolution Test Current 1 Year (23
℃
± 5
℃
)
Resistance
4
100.0000 Ω
100 μΩ
1 mA
0.010 + 0.004
1.000000 kΩ
1 mΩ
1 mA
0.010 + 0.001
10.00000 kΩ
10 mΩ
100 μA
0.010 + 0.001
100.0000 kΩ
100 mΩ
10 μA
0.010 + 0.001
1.000000 MΩ
1 Ω
5 μA
0.010 + 0.001
10.00000 MΩ
10 Ω
500 nA
0.040 + 0.001
100.0000 MΩ
100 Ω
500 nA //10 MΩ
0.800 + 0.010
Diode Test
1.0000 V
10 μV
1 mA
0.010 + 0.020
Continuity
1000.00 Ω
10 mΩ
1 mA
0.010 + 0.030
4.
a. Specifications are for 4-wire ohms. For 2-wire ohms, use Math Null function
b. Max. Lead Resistance: 10% of range per lead for 100Ω and 1KΩ ranges;
1kΩ per lead for all other ranges.
c. Input protection: 1000 V
(PEAK), all ranges.